Macromolecular crystallography beamline BL17U1 at the Shanghai Synchrotron Radiation Facility hasbeenrelocated, upgraded, and given a newID (BL02U1). It now deliversX-rays intheenergyrangeof616 keV, withafocusedbeamof11.6m 4.8 m andphotonfluxgreater than 10^12phs/s. Thehighcredibilityandstabilityofthebeamandgoodtimingsynchronizationof the equipmentsignificantly improvetheexperimentalefficiency. SinceJune2021, whenitofficially openedtousers, over4200 hofbeamtimehavebeenprovidedto over200 researchgroupstocollectdataatthebeamline. Its goodperformanceandstableoperationhave led to the resolution of several structuresbasedondatacollectedatthebeamline.
KeLiuHuanZhouQinXuHuatingKongKunhaoZhangWeiweiWangMinjunLiZhijunWangQiangyanPanXingyaWangFengYuQishengWang
Shanghai Advanced Research Institute, Chinese Academy of SciencesUniversity of Chinese Academy of Sciences
数理科学
Macromolecular crystallography beamline;Shanghai Synchrotron Radiation Facility; BL02U1;
KeLiu,HuanZhou,QinXu,HuatingKong,KunhaoZhang,WeiweiWang,MinjunLi,ZhijunWang,QiangyanPan,XingyaWang,FengYu,QishengWang.BL02U1: The Relocated Macromolecular Crystallography Beamline at the Shanghai Synchrotron Radiation Facility[EB/OL].(2023-11-15)[2023-12-11].http://chinaxiv.org/abs/202311.00134V1.点此复制