纳米ZrO2基电化学发光免疫传感器定量检测水与蔬菜汁中17β-雌二醇OA
Quantitative detection of 17β-estradiol in water and vegetable juice using an electrochemiluminescence immunosensor based on ZrO2 nanoparticles
在玻璃电极表面引入纳米 ZrO2(ZrO2NPs)并固定绵羊源 17β-雌二醇(E2)多克隆抗体(Ab),将此电极浸入 E2 待测溶液中捕获 E2,再引入 E2 标记聚苯胺包裹金纳米颗粒(PANI@AuNPs-BSA-E2),让其结合未与 E2 结合的 Ab,构建出 ZrO2NPs 基 E2 检测电化学发光(ECL)免疫传感器,根据电极 ECL 强度对E2 含量响应实现对待测体系 E2 的定量检测,考察了底物浓度、缓冲溶液环境(pH 与共反应物浓度)、ZrO2NPs悬液浓度、ECL 检测电位扫描速率、Ab 与 E2 作用时间、未与 E2 结合的 Ab 与 PANI@AuNPs-BSA-E2 作用时间等的影响,同时探究了该传感器对 E2 检测的稳定性、选择性和再生性,及其对自然水样和蔬菜汁样的适用性.结果表明:检测传感器最佳孵化与检测条件为:缓冲体系 pH 和共反应物 K2S2O8 浓度分别为 7.0和 100 mmol/L、ZrO2NPs 悬液浓度为 2.5 g/L、Ab 与 E2 作用时间为 2.5 h、未与 E2 结合的 Ab 与PANI@AuNPs-BSA-E2 作用为 2.5 h、电位扫描速率为 0.1 V/s;最佳检测条件下,该传感器的 ECL 强度对 E2浓度(1.0×10-5~100 µg/L)有很好的线性响应,相关系数R2 为 0.998 6,检测限为 0.011 µg/L,同时该传感器对 E2 检测的稳定性和选择性强,再生性好,在自然水样和蔬菜汁样 E2 检测中适用性强.
[Objective]Low-concentration 17β-estradiol(E2)in environmental media(µg/L or even lower)is one of the strongest endocrine disruptors in environmental estrogens.A quantitative analysis of E2 in the environment lacks sufficient sensitivity and accuracy.To fill the gap in existing environmental estrogen detection technology,electrochemiluminescence(ECL)and immunosensor techniques were combined to construct a nano-ZrO2(ZrO2NPs)-based ECL immunosensor(ECLI).The potential of the immunosensor for detecting trace E2 was evaluated.[Methods]First,ZrO2NPs were prepared by a sol-gel method,and polyaniline(PANI)-coated gold nanoparticles(AuNPs)(PANI@AuNPs)were prepared using a gold nano seed coating method.Thereafter,ZrO2NPs were introduced onto the surface of a glass electrode(GCE),and sheep-derived E2 polyclonal antibody(Ab)was immobilized on the electrode to prepare a ZrO2NPs-based ECLI.Finally,the ECLI was immersed in the E2 test solution to capture E2,followed by the introduction of E2-labeled PANI@AuNPs(PANI@AuNPs-BSA-E2)to bind with Ab that was not pre-bound to E2.During this process,the electrochemical impedance spectroscopy(EIS),cyclic voltammetry(CV),and ECL response characteristics of the GCE surface at different stages of incubation were observed.The influence of the following factors on the ECL intensity of the ECLI during E2 detection was simultaneously investigated:substrate concentration,buffer solution environment(pH and co-reactant concentration),concentration of ZrO2NP suspension,potential scanning rate for ECL detection,Ab and E2 interaction time,and interaction time for Ab not bound to E2 with PANI@AuNPs-BSA-E2.E2 in the samples was quantitatively detected based on the relationship between the ECL intensity and E2 concentration.Finally,the stability,selectivity,regenerability,and applicability of the ECLI sensor for E2 detection in natural water and vegetable juice were investigated through 10 consecutive scans.Experiments were carried out by immersing the sensor in the sample for 1-12 days;replacing the target antigen E2 with diethylstilbestrol,estriol,estrone,dichlorodiphenyltrichloroethane,ethinylestradiol,or bisphenol A;adding E2 standard solution;regenerating the sensor via microwave desorption;and immersing it in natural water and vegetable juice samples containing a certain concentration of E2.[Results]The resistance of the multi-level sensor with the surface-modified GCE gradually increased.The ECL signal originated from bonding between Ab and E2 loaded on the surface of the ZrO2NPs.The optimal conditions for E2 detection by the ECLI were as follows:pH 7.0 buffer system;K2S2O8(co-reactant)concentration of 100 mmol/L;ZrO2NPs suspension concentration of 2.5 g/L;Ab and E2;interaction time of 2.5 h for Ab not bound to E2 with PANI@AuNPs-BSA-E2;and potential scanning rate of 0.1 V/s.Under the optimal detection conditions,the ECL intensity of the ECLI exhibited a good linear response to the E2 concentration,with a correlation coefficient(R2)of 0.998 6 and a detection limit of 0.011 µg/L.After 10 consecutive scans and 12 days of storage,the ECL performance of the ECLI still reached≥98%of its original level.The constructed ECLI was not affected by environmental endocrine disruptors.In natural water and vegetable juice,the recovery rate of E2 was in the range of 99.3%-104.0%and 98.7%-103.0%,respectively,indicating the satisfactory detection ability of the sensor.Regeneration using the ultrasound method indicated that ECLI had good regeneration and reusability.[Conclusions]The constructed ZrO2NPs-based ECLI sensor achieved quantitative detection of trace amounts of E2 in environmental samples,with good stability and selectivity,affording a wide detection range,low detection limit,and high sensitivity.
罗哲;袁枫;廖敏;谢晓梅;肖玥;范明笛;樊春林
土壤污染防治与安全全国重点实验室,浙江 杭州 310058||浙江大学 环境与资源学院,浙江 杭州 310058土壤污染防治与安全全国重点实验室,浙江 杭州 310058||浙江大学 环境与资源学院,浙江 杭州 310058土壤污染防治与安全全国重点实验室,浙江 杭州 310058||浙江大学 环境与资源学院,浙江 杭州 310058土壤污染防治与安全全国重点实验室,浙江 杭州 310058||国家级环境与资源实验教学示范中心,浙江 杭州 310058土壤污染防治与安全全国重点实验室,浙江 杭州 310058||浙江大学 环境与资源学院,浙江 杭州 310058土壤污染防治与安全全国重点实验室,浙江 杭州 310058||浙江大学 环境与资源学院,浙江 杭州 310058宁波费尔诺生物科技有限公司,浙江 宁波 315502
信息技术与安全科学
纳米ZrO217β-雌二醇17β-雌二醇多克隆抗体夹心式电化学发光免疫传感器17β-雌二醇检测
nano-ZrO217β-estradiol17β-estradiol polyclonal antibodysandwich electrochemiluminescence immunosensor17β-estradiol detection
《实验技术与管理》 2026 (10)
50-57,8