基于预标定内参的数字体视显微系统标定方法OA
Calibration Method for Digital Stereoscopic Microscopy System Based on Pre-Calibrat-ed Intrinsic Parameters
针对数字体视显微系统浅景深导致通用相机标定方法无法适用的问题,文中基于张正友标定法提出了一种基于预标定内参的数字体视显微系统标定方法.通过变倍率法和显微镜等效模型预标定主点和焦距,精确估计相机内参,利用该内参结合张正友标定法估计和优化其他相机参数,实现了显微环境下对数字体视显微系统的标定.实验结果表明,预标定内参方法能够稳定且精确地获得主点和焦距,为相机参数优化提供合理初值.在实际数字体视显微系统标定中,所提方法在不同常用倍数下均能够准确标定相机参数,重投影误差小于0.5 pixel,光轴夹角为9.6°,接近设备标定参数10°.在利用标定出的相机参数进行深度测量实验时,对50 μm的多次测量平均相对误差为2.29%,均方差为1.5 μm,验证了所提标定方法的有效性与可靠性.
In view of the problem that the shallow depth of field of the digital stereomicroscope system makes the general camera calibration method inapplicable,this study proposes a calibration method for the digital stereomi-croscope system based on the pre-calibrated internal reference based on the Zhang Zhengyou calibration method.By using the variable magnification method and the equivalent model of the microscope to pre-calibrate the princi-pal point and focal length,the internal parameters of the camera are accurately estimated.By combining these inter-nal parameters with Zhang Zhengyou's calibration method to estimate and optimize other camera parameters,the cal-ibration of the logarithmic stereoscopic microscopic system in the microscopic environment is achieved.The experi-mental results show that the pre-calibrated internal reference method can stably and accurately obtain the principal point and focal length,providing reasonable initial values for the optimization of camera parameters.In the calibra-tion of the actual digital stereomicroscope system,the proposed method can accurately calibrate the camera parame-ters under different commonly used magnifications.The reprojection error is less than 0.5 pixel,and the optical ax-is angle is 9.6°,which is close to the equipment calibration parameter 10°.When conducting depth measurement experiments using the calibrated camera parameters,the average relative error of multiple measurements at 50 μm is 2.29%,and the mean square error is 1.5 μm,verifying the effectiveness and reliability of the proposed calibra-tion method.
陈泽;袁嫣红;乐珂
浙江理工大学 机械工程学院,浙江 杭州 310018浙江理工大学 机械工程学院,浙江 杭州 310018浙江理工大学 机械工程学院,浙江 杭州 310018
信息技术与安全科学
显微立体视觉体视显微镜相机标定张正友标定法机器视觉单目标定极线校正
micro stereo visionstereo microscopecamera calibrationZhang Zhengyou calibration meth-odmachine visionsingle targetingepipolar rectification
《电子科技》 2026 (10)
87-94,8
浙江省科技厅项目(2022C01065) Zhejiang Science and Technology Department Project(2022C01065)
评论