O-MAR技术在膝关节单髁置换术后的应用价值OA

The Application Value of O-MAR Technology in Unicompartmental Knee Arthroplasty

中文摘要英文摘要

目的:探讨飞利浦去金属伪影算法(O-MAR)在去除膝关节单髁置换术后金属植入物伪影中的价值.方法:回顾性收集 2024年 1月至 10月在河北医科大学第一医院实施膝关节单髁置换术患者 100例,均应用飞利浦 Incisive 64排 CT进行扫描,运用 O-MAR对扫描后数据进行重建,自动生成 Non O-MAR及 O-MAR两组图像.对比两组重建图像膝关节假体周围骨组织、肌肉的信噪比(SNR)及伪影指数(AI)用以评估伪影情况;采用GE AW4.7工作站对薄层图像数据进行三维重建并分别测量每组数据线束硬化伪影和骨组织及假体体积总和,Non O-MAR组三维图像体积与 O-MAR组三维图像体积之差为伪影减少体积.由两名高年资影像医师对膝关节断层及三维图像进行主观评分用以评估成像质量效果.结果:两组图像 O-MAR组膝关节假体周围骨组织、肌肉 SNR均有升高,AI均有降低,O-MAR组膝关节整体主观评分高于 Non O-MAR组,两者差异具有统计学意义.Non O-MAR组三维图像体积为(259.59±70.76)cm3,O-MAR组三维图像体积为(203.02±58.37)cm3,两者差别有统计学意义.O-MAR组三维图像体积较 Non O-MAR组减少 21.79%.结论:膝关节单髁置换术后患者 CT检查,应用 O-MAR技术可最大限度消除金属伪影,并充分显示邻近组织,更有助于膝关节单髁置换术后复查和临床诊断.

Objective:This study evaluated the value of the Philips Orthopedic Metal Artifact Reduction(O-MAR)algorithm for reducing metal implant phantom following unicompartmental knee arthroplasty.Methods:A retrospective study was conducted on 100 patients who underwent unicompartmental knee arthroplasty at the First Hospital of Hebei Medical University from January to October 2024.All patients underwent CT using a Philips Incisive 64-slice scanner,and the post-scan data were reconstructed using O-MAR to generate images with and without O-MAR correction(O-MAR and non-O-MAR images).The signal-to-noise ratio(SNR)and artifact index(AI)of periarticular bone and muscle tissues in both reconstructed images were compared to evaluate phantom formation.Thin-section images were reconstructed using a GE AW4.7 workstation,followed by measurements of linear streak artifacts and the total volumes of bone tissue and prostheses in each group.The volume reduction due to phantom suppression was calculated as the difference between non-O-MAR and O-MAR volumes.Two senior radiologists conducted subjective evaluations of both cross-sectional and 3D imaging quality.Results:The signal-to-noise ratios(SNRs)of bone and muscle tissues were significantly higher in the O-MAR images than in the non-O-MAR images,whereas the artifact index(AI)was significantly lower in the O-MAR group.The O-MAR group demonstrated significantly higher subjective knee joint scores than the non-O-MAR group.Three-dimensional imaging volumes were measured as(259.59±70.76)cm3 for the non-O-MAR group and(203.02±58.37)cm3 for the O-MAR group,with a statistically significant difference.The three-dimensional image volume in the O-MAR group was 21.79%less than that in the non-O-MAR group.Conclusion:In patients undergoing unicompartmental knee arthroplasty,the O-MAR algorithm effectively reduced metal artifacts,improved image quality,and enhanced visualization of the surrounding tissues.

赵德园;聂关伟;赵君禄;郑立冬;高艺铭

河北医科大学第一医院放射科,石家庄 050031河北医科大学第一医院手术室,石家庄 050031河北医科大学第一医院放射科,石家庄 050031河北医科大学第一医院放射科,石家庄 050031河北医科大学第一医院放射科,石家庄 050031

信息技术与安全科学

植入物膝关节单髁置换术去金属伪影

implantknee-jointunicompartmental knee arthroplastymetal artifact reduction

《CT理论与应用研究》 2026 (6)

1055-1060,6

河北省医学科学研究课题计划(多模态成像技术在膝关节单髁置换术中的诊断价值(20241564)).

10.15953/j.ctta.2025.444

评论