金属伪影抑制算法对能量CT定量测量的固有影响研究OA

Intrinsic Impact of the Metal Artifact Reduction Algorithm on Quantitative Measurement in Spectral CT

中文摘要英文摘要

目的:探讨金属伪影抑制技术 O-MAR对能量 CT图像中不同组织定量参数测量的影响,明确其在无金属背景下的算法固有偏差.方法:使用 CIRS 062M电子密度模体,采用 Philips IQon Spectral CT进行重复扫描(n=10).分析 8种插入物(吸气肺、呼气肺、脂肪、乳腺、肌肉、肝脏、小梁骨和密质骨)在启用与未启用O-MAR算法下的定量结果.重组图像包括常规混合能量图像(Con)、钙抑制图像(CaSupp)(抑制指数分别为25、50和 75)、有效原子序数(Zeff)图及电子密度(ED)图.以未启用 O-MAR的测量值为参照,采用 Bland-Altman分析量化平均偏差及 95%一致性界限,并采用配对 t检验评估统计学差异.结果:O-MAR对定量测量的影响具有组织特异性:① 一致性评价:吸气肺、脂肪及乳腺的一致性较好,偏差较小;而密质骨表现出显著波动,尤其在 CaSupp 75图像上平均偏差达 2.05 HU,95%一致性界限显著宽于其他组织.② 差异性检验:吸气肺、肌肉及脂肪在多数图像类型中差异显著;密质骨在Con及CaSupp 50/75图像中受影响显著;而Zeff 及ED图像表现出最佳的数值稳定性.结论:O-MAR算法在无金属背景下会引入组织特异性的系统性定量偏差,对极低密度组织和高密度骨组织影响显著,但在Zeff 和ED参数中表现稳定.

Objective:To investigate the impact of the orthopedic metal artifact reduction(O-MAR)technique on the measurement of quantitative parameters across various tissues in spectral CT images and define the intrinsic algorithmic bias in a metal-free background.Methods:A CIRS 062M electron density phantom was scanned repeatedly(n=10)using a Philips IQon Spectral CT.Quantitative results for eight tissue-simulating inserts(inspiratory lung,expiratory lung,adipose,breast,muscle,liver,trabecular bone,and dense bone)were analyzed with and without the O-MAR algorithm.Reconstructed images included conventional polyenergetic images(Con),calcium-suppressed(CaSupp)images(with suppression indices of 25,50,and 75),effective atomic number(Zeff)maps,and electron density(ED)maps.Measurements without O-MAR were used as the reference,and Bland–Altman analysis was performed to quantify the mean bias and 95%limits of agreement(95%LoA),supplemented by paired t-tests to evaluate statistical differences.Results:The impact of O-MAR on quantitative measurements exhibited marked tissue specificity:(1)Consistency Evaluation:Inspiratory lung,adipose,and breast tissues demonstrated high consistency with minimal bias.In contrast,the dense bone tissue showed significant fluctuations,particularly in CaSupp 75 images,where the mean bias reached 2.05 HU and 95%LoA was substantially wider than that of other tissues.(2)Difference Testing:Measurements for inspiratory lung,muscle,and adipose tissues showed significant differences across most image types.The dense bone tissue was significantly affected in Con and CaSupp 50/75 images.Conversely,Zeff and ED maps exhibited the highest numerical stability.Conclusion:In a metal-free background,the O-MAR algorithm introduced tissue-specific systematic quantitative biases,with significant impacts on extremely low-density and high-density tissues.However,quantitative parameters in the Zeff and ED maps remained stable.

魏江涛;程雨荷;马梓轩;杨佳硕;张永县;刘丹丹

首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京同仁医院放射科,北京 100730

医药卫生

金属伪影抑制技术能量CT定量测量模体研究

metal artifact reduction technologyspectral CTquantitative measurementphantom study

《CT理论与应用研究》 2026 (6)

1014-1020,7

10.15953/j.ctta.2025.481

评论