首页|期刊导航|中国医学工程|ICU患儿躁动与监护设备假警报的相关性分析及管理对策探讨

ICU患儿躁动与监护设备假警报的相关性分析及管理对策探讨OA

Correlation analysis between agitation in ICU children and false alarms of monitoring equipment and discussion on corresponding management countermeasures

中文摘要英文摘要

目的 探讨ICU患儿躁动与医疗设备假警报的相关性,分析躁动诱发假警报的关联特点.方法 选取2024年1月至2025年6月浙江大学医学院附属儿童医院80例ICU住院患儿为对象,采用改良儿童重症躁动分级量表进行评估,连续记录监护设备警报数据,采用单因素方差分析比较不同躁动程度组间各指标差异,采用Spearman秩相关分析躁动程度与假警报指标的量化关联.结果 患儿总躁动发生率为78.75%(63/80);共记录警报1 264次,假警报896次(假警报率70.89%),其中躁动相关假警报379次,占全部假警报的42.30%;不同躁动分级患儿躁动相关假警报次数差异有统计学意义(P=0.001);Spearman秩相关分析显示,躁动程度与躁动相关假警报呈中等正相关(rs=0.445,P<0.001),躁动程度与日均假警报次数无明显相关性(rs=0.015,P=0.896).结论 ICU患儿躁动与监护设备假警报存在关联,躁动相关假警报是一个可独立识别、精准干预的靶点;结合躁动与假警报的量化关联规律,本文提出躁动分层精准管理思路,可为儿科重症假警报防控提供理论参考.

[Objective]To explore the correlation between agitation in ICU children and false alarms of medical equipment,and analyze the association characteristics of false alarms induced by agitation.[Methods]Eighty children admitted to the pediatric ICU from January 2024 to June 2025 were enrolled,and agitation was graded by a modified pediatric critical agitation scale.Alarm data of monitoring equipment were continuously recorded.One-way ANOVA was used to compare the differences in various indicators among groups with different agitation grades,and Spearman rank correlation analysis was used to analyze the quantitative association between agitation severity and false alarm indicators.[Results]The overall incidence of agitation in children was 78.75%.A total of 1,264 alarms were recorded,including 896 false alarms(false alarm rate 70.89%),of which 379 were agitation-related false alarms,accounting for 42.30%of all false alarms.There was a statistically significant difference in the number of agitation-related false alarms among children with different agitation grades(P=0.001).Spearman rank correlation analysis showed that agitation severity was moderately and significantly positively correlated with agitation-related false alarms(rs=0.445,P<0.001),while there was no significant correlation between agitation severity and average daily false alarm frequency(rs=0.015,P=0.896).[Conclusion]There is an association between agitation in ICU children and false alarms of monitoring equipment.Agitation-related false alarms are independently identifiable and precisely intervenable targets.Combined with the quantitative association between agitation and false alarms,this paper proposes a grading-based precise management strategy for agitation,which can provide theoretical reference for the prevention and control of false alarms in pediatric critical care.

林忠款;周军;马铉超;郑彩仙;郑焜

浙江大学医学院附属儿童医院 医疗设备部,浙江 杭州 310052浙江大学医学院附属儿童医院 CICU,浙江 杭州 310052浙江大学医学院附属儿童医院 医疗设备部,浙江 杭州 310052浙江大学医学院附属儿童医院 医疗设备部,浙江 杭州 310052浙江大学医学院附属儿童医院 医疗设备部,浙江 杭州 310052

医药卫生

儿童躁动医疗设备假警报警报疲劳

childrenagitationmedical equipmentfalse alarmalarm fatigue

《中国医学工程》 2026 (8)

1-4,4

浙江省医药卫生科技计划项目(2023KY831)中国医学装备协会医疗装备综合价值评估研究项目(ZGYXYP2025209)

10.19338/j.issn.1672-2019.2026.08.001

评论