磁共振加速器液氦量变化及其剂量学影响分析OA
Analysis on changes of liquid helium volume of MR-linac and its dosimetric influence
目的:分析磁共振加速器冷头异常停机至恢复全流程中磁体压力与液氦量的动态变化,探究不同液氦水平下射束穿过低温恒温器的透射表征(CTC)曲线差异,明确液氦状态对束流衰减及机架角度相关透射特性的影响规律;为磁共振加速器的输出校准修正与治疗计划系统(TPS)中束流特性的精准表征提供参考.方法:通过医科达Unity磁共振直线加速器扫描工作站的压力和液氦容量感应日志文件,分析随着冷头停机后,磁体的压力及液氦量的变化与故障持续时间的关系;冷头恢复运转96 h后,采用CTC专用测量平台,将配有水填充建成帽的PTW31013电离室架置于等中心处,测量不同机架角度射线穿过低温恒温器后的剂量衰减并归一至0°,并与装机时刻的测量结果进行比较.结果:冷头停止工作后的48 h内,磁体压力从206.8 kPa的基线值升高至2 194 kPa,呈前快后慢趋势,液氦量累计减少3.4%;冷头恢复工作后,压力值呈线性降低,经过96 h后恢复至基线水平.液氦损失前后,CTC测量对比结果显示,以机架角度0°和180°为界,在0°~180°和180°~360°范围之间剂量衰减变化存在差异,平均值分别为0.52%和-0.48%,其中最大偏差为0.86%和-1.65%,出现的机架角度分别为76°和294°.结论:Unity磁共振加速器磁体在冷头停机后压力升高、液氦量减少,存在失超风险,应及时发现并消除故障,确保设备正常运行.液氦损失导致剂量学影响可>1%,应通过重新测量CTC曲线加以修正.
Objective:To explore the differences of Cryostat Transmission Characterization(CTC)curves that the beam passes through cryostat under different liquid helium volumes,and confirm the influence law of liquid helium state on beam attenuation and gantry-angle-related transmission characteristics by analyzing the dynamic changes of magnet pressure and liquid helium volume of MR-linac cold head during the full process from abnormal shutdown to recovery,so as to provide references for output calibration correction of MR-linacs,and accurate characterization of beam properties in treatment planning system(TPS).Methods:The relationships of the change of pressure and liquid helium volume of magnetic body,and duration of fault were analyzed after cold head was closed down through log files about sensing of pressure and liquid helium volume of scanning workstation of Elekta Unity MR-linac accelerator.After 96 h that cold head recovered operation,the CTC special measure platform was adopted to take PTW31013 ionization chamber with a water-filled build-up cap at the isocenter.The dose attenuation was measured after the ray of different gantry angle passed through the cryostat,and the results were normalized to 0°,which were compared with the measured results at the time of installation.Results:Within 48 h after cold head was shut down,the pressure of magnet body increased from the 206.8 kPa of baseline to 2 194 kPa,which appeared a trend from fast in the front to slow in the back,in addition,the liquid helium volume cumulatively reduced 3.4%.After cold head recovery,the pressure appeared linear decrease,and it recovered to baseline level after 96 h.The comparative results of CTC measures before and after liquid helium loss showed the change of dose attenuation existed differences between range of 0°-180°,and between range of 180°-360° when the 0° and 180° of gantry angles were used as boundary value,and the average values were respectively 0.52%and-0.48%.The maximum deviations of them were respectively 0.86%and-1.65%,and the occurred gantry angles were respectively 76° and 294°.Conclusion:The pressure of magnet body of Unity MR-linac enhances,and the liquid helium volume of that reduces after the cold head shut down,which exists risk of losing superconductivity.It should be timely found,and the fault of that should be timely eliminated,so as to ensure normal operation of equipment.The dosimetric influence that is caused by loss of liquid helium can exceed 1%,and it should be corrected through measures CTC curve again.
孙玉亮;管秋;张福泉;徐浩然;李洪明;王冠群;于浪
中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730中国医学科学院 北京协和医学院 北京协和医院放射治疗科 北京 100730
医药卫生
磁共振加速器冷头停机磁体压力液氦量低温恒温器的透射表征
Magnetic resonance linac(MR-linac)Cold head shutdownPressure of magnet bodyLiquid helium volumeTransmission characterization of cryostat
《中国医学装备》 2026 (9)
14-17,4
国家市场监督管理总局科技计划项目(2024MK196)国家重点研发计划-国产创新一体化放疗设备的技术解决方案研究(2022YFC2407101) Science and Technology Program Project of the State Administration for Market Regulation(2024MK196)National Key Research and Development Program-Technical Solution Research for Domestic Innovative Integrated Radiotherapy Equipment(2022YFC2407101)