同步视觉型人工晶体的阳性视觉干扰评价方法研究OA
Evaluation method for positive dysphotopsia of simultaneous vision intraocular lenses
目的:为定量评估人工晶体(intraocular lens,IOL)光学设计引入的不良视觉干扰风险,提出一种基于阳性视觉干扰敏感因子(positive dysphotopsia sensitive factor,PDSF)的评价方法.方法:选取5款标称光焦度为20.0 D、焦点不同的IOL样品,依据临床功能和光学设计类型分为对照组(单焦点,非球面折射设计)、试验1组(双焦点,非球面衍射设计)、试验2组(三焦点,非球面衍射设计)、试验3-1组(焦深扩展,折射与衍射混合设计)和试验3-2组(焦深扩展,非球面衍射设计).采用OptiSphere® IOL R&D全自动焦距仪进行光学性能测试,通过测算远距焦面0 D、远至中距过渡焦面(0.5±0.5)D、中距焦面(1.5±0.5)D、近距焦面(2.5±0.5)D 4个焦面位置的斯特列尔比率(Strehl ratio,SR)、光斑直径(spot diameter,SD)及周边光能占比(peripheral energy ratio,PER)3项指标,计算PDSF值,量化阳性视觉干扰相关的光能损失、光斑扩散和周边光能分布情况.结果:对照组远距焦面的PDSF为0.33,在远至中距过渡焦面升至2.14.试验1组和试验2组远距焦面处的PDSF分别为0.52和0.68,均高于对照组,光晕和眩光现象比对照组明显;离焦后试验1组和试验2组PDSF分别升至1.26和1.09,在中距焦面位置光能叠加效应显著,视觉干扰严重.试验3-1组远距焦面PDSF为0.36,中距焦面PDSF为0.67.试验3-2组远距焦面PDSF为0.75,中距焦面PDSF为0.66,离焦后PDSF保持稳定.结论:提出的PDSF评价方法实现了IOL不良视觉干扰效应的系统性量化评估,初步验证了PDSF与IOL临床研究中的阳性视觉干扰发生的相关性,为IOL术前优化设计及风险预测提供了新的量化工具.
Objective To propose an evaluation method based on the positive dysphotopsia sensitive factor(PDSF)for quan-titatively assessing the risk of adverse visual disturbances introduced by intraocular lens(IOL)optical designs.Methods Five IOL samples with nominal optical focal powers of 20.0 D and different focal points were selected and classified into a control group with monofocal and aspherical refractive design,Test Group 1 with bifocal and aspherical diffractive design,Test Group 2 with trifocal and aspherical diffractive design,Test Group 3-1 with depth-of-focus extended and hybrid refractive-diffractive design and Test Group 3-2 with depth-of-focus extended and aspherical diffractive design.Optical performance testing was performed using an OptiSphere® IOL R&D fully automated focal length meter.PDSF was calculated by measuring three indexes of Strehl ratio(SR),spot diameter(SD)and peripheral energy ratio(PER)for the four focal positions of 0 D at the far focal plane,(0.5±0.5)D at the far-to-mid focal plane,(1.5±0.5)D at the mid focal plane and(2.5±0.5)D at the near focal plane,so as to quantify the light energy loss,spot diffusion and peripheral light energy distribution associated with positive visual disturbance.Results The control group had PDSF being 0.33 and rising to 2.14 at the far-to-mid focal plane.Test Group 1 and 2 had PDSF at the far focal plane being 0.52 and 0.68 respectively,both higher than that in the con-trol group,which had more obvious halo and glare when compared with the control group;in Test Group 1 and 2 the values of PDSF increased to 1.26 and 1.09 respectively after defocusing,with significant light-energy superposition effect and serious visual disturbance at the mid focal plane.Test Group 3-1 had PDSF being 0.36 after defocusing at the far focal plane and 0.67 at the mid focal plane.In Test Group 3-2 PDSF was 0.75 and 0.66 respectively at the far and mid focal planes,which remained stable after defocusing.Conclusion The proposed PDSF evaluation method can be used for systematic quantitative assessment of the adverse visual disturbance effects of IOL,validating preliminarily the correlation between PDSF and the occurrence of positive visual disturbance in IOL clinical studies and providing a new quantitative tool for the optimization of IOL preoperative design and risk prediction.[Chinese Medical Equipment Journal,2026,47(9):32-40]
郝琦;冯勤;宋婷;傅敏;陈琼慧
浙江省医疗器械检验研究院,杭州 310018浙江省医疗器械检验研究院,杭州 310018国家药品监督管理局生物医学光学重点实验室,杭州 310018国家药品监督管理局生物医学光学重点实验室,杭州 310018浙江省医疗器械安全性评价研究重点实验室,杭州 310018
医药卫生
同步视觉型人工晶体阳性视觉干扰敏感因子阳性视觉干扰斯特列尔比率光斑直径周边光能占比光学设计
simultaneous vision intraocular lenspositive dysphotopsia sensitive factorpositive dysphotopsiaStrehl ratiospot diameterperipheral energy ratiooptical design
《医疗卫生装备》 2026 (9)
32-40,9
省属科研院所扶持经费项目(20230001)
评论