针对心脏起搏器固定钩界面整合差与易感染的临床痛点,本文采用微弧氧化技术在纯钛表面制备了镁掺杂的多孔 TiO2 涂层.结果表明,适量镁掺杂未改变锐钛矿相,但显著均化了多孔拓扑形貌,亲水性大幅提升(接触角低至 31.2°).改性涂层展现出强效广谱抑菌活性(抑菌率超90%).其中,TiO2-Mg2 组综合性能最优:溶血率仅 1.30%,自由基清除率达 86.2%,细胞存活率达 95.5%且显著促进心肌细胞铺展.该涂层通过物理-化学协同调控,兼顾了促组织整合、抗凝血与抗氧化功能,为心血管植入器械表面功能化提供了坚实参考.

To address the clinical challenges of poor interfacial integration and high infection risk associated with pacemaker fixation hooks,magnesium-doped porous TiO2 coatings were prepared on pure titanium surfaces via micro-arc oxidation.The results indicate that appropriate magnesium doping did not alter the anatase phase but significantly homogenized the porous topological morphology,leading to a substantial improvement in hydrophilicity(with the contact angle reduced to 31.2°).The modified coatings exhibite potent broad-spectrum antibacterial activity(inhibition rate>90%).Notably,the TiO2-Mg2 group demonstrate the optimal comprehensive performance:a hemolysis rate of only 1.30%,a free radical scavenging rate of 86.2%,and a cell viability of 95.5%,while significantly promoting the spreading of cardiomyocytes.Through synergistic physical-chemical regulation,this coating achieves a combination of enhanced tissue integration,favorable hemocompatibility,and antioxidation functions,providing a solid reference for the surface functionalization of cardiovascular implant devices.

巨鹏亮;奚正星;赵金木;朱永都;马凤仓

上海理工大学 东方泛血管学院,上海 200093上海理工大学 材料与化学学院,上海 200093上海理工大学 材料与化学学院,上海 200093上海理工大学 材料与化学学院,上海 200093上海理工大学 材料与化学学院,上海 200093

矿业与冶金

微弧氧化镁掺杂抗菌性能生物相容性心血管植入器械

micro-arc oxidationMagnesium dopingantibacterial propertybiocompatibilitycardiovascular implants

《有色金属材料与工程》 2026 (3)

39-48,10

国家自然科学基金资助项目(51771119)上海市自然科学基金资助项目(17ZR1419600)