目的:立体定向放射外科是颅内、脊柱、胸腹部恶性肿瘤精准放射治疗的核心手段,摆位误差会直接降低肿瘤靶区照射剂量、加重正常组织放射性损伤.本研究检测分析 CyberKnife M6 立体定向放射外科治疗系统的定位精度以及实施位置误差校正技术的准确性,为各类恶性肿瘤立体定向放射治疗质控提供依据.方法:检测 Cy-berKnife M6 立体定向放射外科治疗系统中图像引导系统自动配准的稳定性和重复性:应用立体定向剂量验证模体(stereotactic dose verification phantom-REF 91250)模拟患者在治疗过程中产生的摆位误差,测量并分析该系统在 X(上/下方向)、Y(左/右方向)、Z(前/后方向)三个线性方向和 r(前后旋转方向)、p(螺距旋转方向)两个旋转方向的定位精度,再用立体定向放射外科(stereotactic radiosurgery,SRS)Mapcheck 模体检测 CyberKnife M6 SRS 治疗系统在实施位置补偿照射技术过程中剂量实施的准确性.结果:CyberKnife M6 SRS 治疗系统中图像引导系统总平均定位精度在线性和旋转方向上的重复性偏差分别≤0.2 mm 和≤0.3°,说明该系统有较好的机械稳定性;同时 CyberKnife M6 系统在实施位置补偿照射技术的剂量准确性为 100%.结论:CyberKnife M6在放射治疗图像采集、自动配准阶段具备优异机械稳定性与重复性,针对恶性肿瘤放射治疗的线性、旋转维度位置误差校准精度高;肿瘤治疗过程中应用位置误差校正技术可保证肿瘤靶区剂量精准投递,减少正常组织受照损伤,满足颅内、胸腹部、脊柱等各类恶性肿瘤立体定向放射治疗临床使用要求.

Objective:Stereotactic radiosurgery(SRS)is the cornerstone of precision radiotherapy for intracranial,spi-nal,and thoracoabdominal malignancies.However,positioning errors may cause target misalignment,resulting in underdos-age of the tumor volume and excessive irradiation of normal tissues.This study evaluated the positioning accuracy of the Cy-berKnife M6 SRS system and the efficacy of its position error correction technology,providing a basis for quality control in stereotactic radiotherapy(SRT)for various malignancies.Methods:The study assessed the stability and reproducibility of image-guided automatic registration in the CyberKnife M6 system.Using the stereotactic dose verification phantom(REF 91250),simulated positioning errors during treatment were created to measure and analyze the system's positioning accuracy across three linear axes[X(superior-inferior),Y(left-right),Z(anterior-posterior)]and two rotational axes(r-axis for roll,p-axis for pitch).The SRS Mapcheck phantom was then employed to verify dose delivery accuracy during position compensation irradiation.Results:The CyberKnife M6 system demonstrated excellent mechanical stability,with the total average positioning accuracy exhibiting repeatability devia-tions of≤0.2 mm along the linear axes and≤0.3° along the rotational axes.Notably,the CyberKnife M6 system demon-strated 100%dose delivery accuracy during irradiation with the position correction technique enabled.Conclusion:The Cy-berKnife M6 system demonstrates excellent mechanical stability and high precision in position error correction during image-guided radiotherapy.This correction technique ensures accurate target dose delivery while sparing normal tissues,fully sup-porting its clinical application in SRS for intracranial,thoracic,abdominal,and spinal malignancies.

黄燕梅;廖雄飞;周沫涵;高绪峰;应微

610041 成都,四川省肿瘤医院·研究所,四川省肿瘤临床医学研究中心,四川省癌症防治中心,电子科技大学附属肿瘤医院 放疗科610041 成都,四川省肿瘤医院·研究所,四川省肿瘤临床医学研究中心,四川省癌症防治中心,电子科技大学附属肿瘤医院 放疗科610041 成都,四川省肿瘤医院·研究所,四川省肿瘤临床医学研究中心,四川省癌症防治中心,电子科技大学附属肿瘤医院 放疗科610041 成都,四川省肿瘤医院·研究所,四川省肿瘤临床医学研究中心,四川省癌症防治中心,电子科技大学附属肿瘤医院 放疗科610041 成都,四川省肿瘤医院·研究所,四川省肿瘤临床医学研究中心,四川省癌症防治中心,电子科技大学附属肿瘤医院 放疗科

医药卫生

射波刀定位精度位置误差校正技术自动配准图像引导放射治疗

CyberKnifePositional accuracyPosition correction techniqueAutomatic registrationImage-guided radio-therapy

《肿瘤预防与治疗》 2026 (9)

785-790,6

10.3969/j.issn.1674-0904.2026.09.008

评论