零气发生器性能能否满足标准要求,直接影响后续分析结果的准确性.为了检测零气发生器性能符合国家标准GB 18285-2018要求,设计了一种使用傅里叶变换红外气体分析仪检测零气发生器性能的方法.通过动态稀释法将不同种类的国家有证标准气体稀释到国标要求的检测用特定浓度值,文中称其为"参考标准气体".傅里叶变换红外气体分析仪测量参考标准气体的响应值,并考察分析仪对不同参考标准气体的检出限.依次将国家有证标准气体先后输入A、B、C 3个厂家零气发生器,输出气体使用傅里叶变换红外气体分析仪测量响应值,并与参考标准气体的响应值做除法.结果表明,3个厂家零气发生器都能满足国标的要求,该方法在零气发生器校准检测中的良好应用表现,提高了零气发生器检测效率和可靠性,为零气质量检测提供重要参考.

The performance of a zero-gas generator directly affects the accuracy of the subsequent analysis results.To determine whether the performance meets the requirements of the national standard GB 18285-2018,a detection method incorporating a Fourier-transform infrared(FT-IR)gas analyzer was designed.Various nationally certified standard gases were dynamically diluted to the specific concentrations required by the national standard for detection,which are referred to as"reference standard gases"in the text.The FT-IR gas analyzer measured the responses of the reference standard gases,and the detection limits of the analyzer were determined for various reference standard gases.National certified standard gases were successively input into zero-gas generators from three different manufacturers:A,B,and C.The responses of the output gases were measured using the FT-IR gas analyzer and then divided by the responses of the reference standard gases.The results showed that the zero-gas generators from all three manufacturers met the requirements of the national standard.The robust performance of this method for the calibration and assessment of detection capability of zero-gas generators improves their detection efficiency and reliability,providing an important reference for the quality detection of zero gas.

吕利;夏春;李剑;程丽萍;王海燕;杨倩;王国娟

青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101青岛市计量技术研究院,山东青岛 266101

化学化工

零气发生器标准物质傅里叶变换红外气体分析仪动态稀释法检出限

zero air generatorreference materialFT-IR spectroscopydynamic dilution methoddetection limit

《化学试剂》 2026 (5)

48-53,6

山东计量测试学会科技项目(2024KJ31).

10.13822/j.cnki.hxsj.2026.0009

评论