X射线能量对剂量计测量CT辐射剂量的影响OA
The Effect of X-ray Energy on Dosimeter Measurements of CT Radiation Dose
目的:使用Raysafe-X2 空气电离室(简称"RX2")、piranha CT dose profiler半导体电离室(简称"CDP")和30013 Farmer空气电离室(简称"PT3")测量不同X射线能量下的点剂量和容积CT剂量指数(CTDIvol),探讨X射线能量对不同剂量计评估辐射剂量的影响.方法:逐层扫描管电压设置,单能量模式包括:80、90、100、120 和 140kV 以及150kV(Sn),双能量模式包括:70/150kV(Sn)、80/150kV(Sn)、90/150kV(Sn)、100/150kV(Sn)和80/140kV.通过手动调节管电流,保证各组CTDIvol约为20mGy,记录显示CTDIvol与剂量长度乘积(DLP).使用CDP和PT3 测量32cm模体中心、0、3、6 和9 点位点剂量值.使用RX2、CDP和PT3测量并计算实际CTDIvol.结果:①点剂量:单能量模式(除Sn 150kV)时CDP在中心位置测得点剂量高于PT3.低能量时非中心位置,CDP测得的75%点剂量高于PT3;高能量时PT3 测得的88%的点剂量高于CDP.双能量模式(除90+Sn 150kV)时CDP在中心位置测得点剂量均高于PT3.除80+140kV,非中心位置PT3测得点剂量高于CDP.②CTDIvol:RX2 测量值为(18.89±0.38)mGy,CDP测量值为18.31(19.25,20.84)mGy,PT3测量值为(20.35±0.38)mGy.除使用CDP在150kV(Sn)的射线能量下,余射线能量下3 种剂量计测得CTDIvol均在16~24mGy之间.结论:不同剂量计对射线能量的响应特性存在差异,因此在不同测量条件和测量目的下,选择合适的剂量计至关重要.
Objective:The point dose and volume CT dosimetry index(CTDIvol)at different X-ray energies were measured using the Raysafe-X2 air ionization chamber(referred to as"RX2"),the Piranha CT dose profiler semiconductor ionization chamber(referred to as"CDP"),and the 30013 Farmer air ionization chamber(referred to as"PT3"),in order to explore the effect of X-ray energy on the radiation dose evaluation by different dose meters.Methods:The tube voltage settings for each layer were as follows:single energy mode included 80kV,90kV,100kV,120kV,140kV,and 150kV(Sn);dual energy mode included 70/150kV(Sn),80/150kV(Sn),90/150kV(Sn),100/150kV(Sn),and 80/140kV.The tube current was manually adjusted to ensure that the CTDIvol for each group was approximately 20 mGy,and the displayed CTDIvol and DLP were recorded.Point dose values were measured at the center,0-point,3-point,6-point,and 9-point positions of the 32cm phantom using CDP and PT3.The actual CTDIvol was measured and calculated using RX2,CDP,and PT3.Results:(1)Point Dose:In single energy mode(except for Sn150kV),CDP measured higher point doses at the center compared to PT3.At low energies,the point dose measured by CDP at non-center positions was 75%higher than that of PT3,while at high energies,88%of the point doses measured by PT3 were higher than those measured by CDP.In dual energy mode(except for 90+Sn150kV),CDP measured higher point doses at the center compared to PT3.However,except for 80+140kV,PT3 measured higher point doses than CDP at non-center positions.(2)As for CTDIvol,the measured values were as follows:RX2 measured(18.89±0.38)mGy,CDP measured 18.31(19.25,20.84)mGy,and PT3 measured(20.35±0.38)mGy.Except when CDP was used at 150kV(Sn),the CTDIvol measured by all three dose meters for other X-ray energies ranged between 16~24mGy.Conclusion:Different dosimeters exhibit varying responses to X-ray energy.Therefore,selecting an appropriate dosimeter is crucial depending on the measurement conditions and objectives.
彭乐欣;马梓轩;牛延涛;张永县;胡玲静;吴宁;高行;刘丹丹
首都医科大学附属北京同仁医院放射科,北京 100730||首都医科大学燕京医学院,北京 101300首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学附属北京友谊医院放射科,北京 100050首都医科大学附属北京同仁医院放射科,北京 100730首都医科大学燕京医学院,北京 101300首都医科大学燕京医学院,北京 101300山东第一医科大学(山东省医学科学院)放射学院,山东 泰安 271016首都医科大学附属北京同仁医院放射科,北京 100730
医药卫生
X射线能量剂量计点剂量容积CT剂量指数
X-ray energydosimeterpoint dosevolume CT dosimetry index
《CT理论与应用研究》 2026 (3)
448-453,6
北京市教育科学"十四五"规划2024年度一般课题(CDDB24223).
评论