金宝AK98血液透析机故障维修3例OA
Fault maintenance of Gambro AK98 Hemodialysis Equipment:Three case reports
介绍了金宝AK98血液透析机开机自检后频繁出现的电导率超出限制、漏血探测器测量错误、开机自检报"静脉夹测试失败,无法打开"3例故障,分析了故障成因,提出了具体的排除方法,为设备维修人员维修同类故障提供了参考.
Three cases of faults of Gambro AK98 Hemodialysis Equipment were explored in terms of phenomena,causes and maintenance measures,including conductivity exceeding limits frequently after power-on self test,malfunction of blood leakage detector and error message stating"Vein clip test failed,cannot open"during power-on self test.References were provided for maintenance technicians to treat similar faults.[Chinese Medical Equipment Journal,2026,47(3):118-120]
潘方娣
广州中医药大学顺德医院(佛山市顺德区中医院)设备科,广东 佛山 528300
医药卫生
金宝AK98血液透析机故障维修故障排除
Gambro AK98 Hemodialysis Equipmentfault maintenancetroubleshooting
《医疗卫生装备》 2026 (3)
118-120,3
评论