金宝AK98血液透析机故障维修3例OA

Fault maintenance of Gambro AK98 Hemodialysis Equipment:Three case reports

中文摘要英文摘要

介绍了金宝AK98血液透析机开机自检后频繁出现的电导率超出限制、漏血探测器测量错误、开机自检报"静脉夹测试失败,无法打开"3例故障,分析了故障成因,提出了具体的排除方法,为设备维修人员维修同类故障提供了参考.

Three cases of faults of Gambro AK98 Hemodialysis Equipment were explored in terms of phenomena,causes and maintenance measures,including conductivity exceeding limits frequently after power-on self test,malfunction of blood leakage detector and error message stating"Vein clip test failed,cannot open"during power-on self test.References were provided for maintenance technicians to treat similar faults.[Chinese Medical Equipment Journal,2026,47(3):118-120]