基于镓熔点新内插方法可行性探索OA
Feasibility Study of a Novel Interpolation Method Based on Gallium Melting Point
在不改变90国际温标(ITS-90)的内插方程和参考函数的前提下,探索将镓熔点代替水三相点作为ITS-90电阻比参考点的可行性.针对广泛使用的273.16~933.473 K和83.808 3~273.16 K两大温区,分别选取37支和27支标准铂电阻温度计的校准数据,采用镓熔点作为电阻比参考点建立新内插方程,并与基于水三相点原ITS-90内插方程进行比较.结果表明:在273.16~933.473 K温区内将电阻比参考点由水三相点替换为镓熔点后,与原内插方程结果的偏差仅为11 μK,在83.808 3~273.16 K范围内,偏差不超过0.5 mK;通过不确定度传播分析,新旧两个方程偏差分别在0.4 mK范围内.这一方法可为ITS-90的应用提供了新的视角和实用性.
To investigate the feasibility of substituting the triple point of water with the gallium melting point as the resistance ratio reference point for the International Temperature Scale of 1990(ITS-90),the interpolation equations and reference functions defined by ITS-90 were preserved unchanged.Calibration data were obtained from 37 standard platinum resistance thermometers for the temperature range of 273.16 K to 933.473 K and from 27 thermometers for the range of 83.808 3 K to 273.16 K.A novel interpolation equation was developed using the gallium melting point as the reference point and subsequently compared with the original ITS-90 equation based on the triple point of water.Results revealed that in 273.16~933.473 K,the deviation from the original equation after substitution was merely 11 μK,while in 83.808 3~273.16 K,the deviation did not exceed 0.5 mK.Uncertainty propagation analysis indicated that deviations of both equations were within 0.4 mK.The approach provides a novel perspective and enhanced feasibility for the application of the ITS-90.
陈松奎;李婷;孙建平;王光耀;胡靖;毋愚;李嘉豪;陶雨佳
中国计量科学研究院,北京 100029中国计量科学研究院,北京 100029||红外遥感定标与计量技术创新中心,北京 100088中国计量科学研究院,北京 100029||红外遥感定标与计量技术创新中心,北京 100088中国计量科学研究院,北京 100029||红外遥感定标与计量技术创新中心,北京 100088中国计量科学研究院,北京 100029中国计量科学研究院,北京 100029中国石油大学,北京 102249中国石油大学,北京 102249
通用工业技术
温度计量90国际温标镓熔点水三相点内插方程不确定度传播
thermometerITS-90gallium melting pointtriple point of waterinterpolation equationuncertainty propagation
《计量学报》 2026 (2)
159-165,7
国家自然科学青年科学基金(52306121)中国计量科学研究院基本科研业务费重点领域项目(AKYZD2301-1)
评论