基于光谱CT在低剂量对比剂和流率肺动脉成像中的可行性研究OA
Feasibility Study of Spectral CT for Pulmonary Angiography Using Low-volume Contrast Agent and Low-flow-rate Protocol
目的:探讨低对比剂剂量和流率在双层探测器光谱CT进行肺动脉CTA(CTPA)检查的可行性.方法:收集 80例临床疑似肺动脉栓塞行CTPA检查的患者,随机分为两组,每组各 40例,低剂量组对比剂剂量15 mL,流率 2.0 mL/s,所得图像行 40 keV单能谱重建;常规组使用小剂量测试法,对比剂总剂量 50 mL,流率5.0 mL/s.记录并计算两组患者图像的容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效剂量(ED)、CT值、信噪比(SNR)和对比噪声比(CNR),并对图像质量进行主观评分.结果:① 常规组和低剂量组的肺动脉干CT值分别为368.450 HU和569.050 HU,差异具有统计学意义;两组间SNR、CNR差异具有统计学意义;② 两组图像质量的主观评分,差异无统计学意义;③ 常规组和低剂量组的DLP分别为 376.500 mGy·cm和294.500 mGy·cm;CTDIvol分别为 11.1502 mGy和 8.350 mGy,ED分别为 5.271 mSv和 4.123 mSv,差异具有统计学意义.结论:双层探测器光谱CT在肺动脉CTA检查中应用低对比剂剂量和流率的扫描方案可以满足诊断需求,降低对比剂渗漏和对比剂肾病等风险.
Objective:To explore the feasibility of using a low-dose contrast agent volume and a low flow rate in dual-layer detector spectral computed tomography(CT)for CT pulmonary angiography(CTPA).Methods:In total,80 patients clinically diagnosed with pulmonary embolisms were collected for CTPA examination and randomly divided into two groups,comprising 40 cases each.The low-dose group received a contrast agent dose of 15 mL at a flow rate of 2 mL/s,and the images were reconstructed using a 40 keV monochromatic spectrum.The conventional group underwent a small-dose test with a contrast agent dose of 50 mL at a flow rate of 5.0 mL/s.The volume CT dose index(CTDIvol),dose-length product(DLP),effective dose(ED),CT values,signal-to-noise ratio(SNR),and contrast-to-noise ratio(CNR)were recorded and calculated for both groups,and the image quality was subjectively scored.Results:(1)The differences in these indicators between the two groups were statistically significant.The pulmonary artery CT values for the conventional and the low-dose groups were 368.450 HU and 569.050 HU,respectively,with a statistically significant difference between the two groups.There were statistically significant differences in SNR and CNR between the two groups.(2)There was no statistically significant difference in the subjective scoring of image quality between the two groups.(3)The DLP values for the conventional and low-dose groups were 376.500 mGy/cm and 294.500 mGy/cm,respectively;the CTDIvol values were 11.150 mGy and 8.350 mGy,respectively;and the ED values were 5.271 mSv and 4.123 mSv,respectively,with statistically significant differences between the groups.Conclusion:The scanning protocol with the"low"contrast agent dose and flow rate using dual-layer detector spectral CT for CTPA examinations met the diagnostic requirements while reducing risks such as contrast extravasation and contrast-induced nephropathy.
朱思思;谢孟臻;漆安琪;罗佳鑫;吴斌;杨忠现;刘于宝
南方医科大学深圳医院医学影像中心,广东 深圳 518110南方医科大学深圳医院医学影像中心,广东 深圳 518110南方医科大学深圳医院医学影像中心,广东 深圳 518110浙江中医药大学第四临床医学院,杭州 310059南方医科大学深圳医院医学影像中心,广东 深圳 518110南方医科大学深圳医院医学影像中心,广东 深圳 518110南方医科大学深圳医院医学影像中心,广东 深圳 518110
医药卫生
双层探测器光谱CT肺动脉CTA低剂量对比剂单能谱成像
dual-layer detector spectral CTpulmonary artery CTAlow-dose contrast agentmonochromatic imaging
《CT理论与应用研究》 2026 (2)
231-238,8
评论