冷阴极牙科低剂量三维成像设备研制OA

Development of a Low-dose 3D Dental Imaging Device Based on Cold Cathodes

中文摘要英文摘要

研制了一种基于碳纳米管(Carbon Nanotube,CNT)冷阴极分布式X射线源和高帧率口内探测器的牙科低剂量三维成像设备,通过电子控制设备等部件和层析重建算法的详细设计,利用 7 个焦点冷阴极X射线管的快脉冲多角度成像特性,实现了低剂量、高清、三维成像目标.该设备阳极电压为 70 kV,阳极电流为 7 mA,各焦点轮换工作,在1s内快速获取 7 个不同角度的图像,并重建出可分层显示的牙科组织的三维高清图像,具有比二维图像更丰富的细节信息.测试表明,阳极电压实测值为 69.6 kV,铝的半值层厚度为 2.13 mm,适合牙齿诊断;设备的线对卡分辨率达到 16 lp·mm-1,低对比分辨率测试可以清晰看出直径为 0.75 mm的孔,具有高清成像特点;在三维成像模式,各焦点工作阳极电流为 7 mA,7 个焦点总成像扫描时间约为 700 ms,快速成像便于临床操作;单次三维成像的空气比释动能面积乘积为 46.4 mGy·cm2,远低于常规锥形束计算机断层扫描(Cone Beam Computed Tomography,CBCT)的照射剂量.三维成像测试观测到了牙齿隐裂现象,具有较好的临床应用价值.

A low-dose 3D dental imaging system based on carbon nanotube(CNT)cold cathode multi-beam X-ray sources and a high-frame-rate intraoral detector is developed.By integrating components such as electronic control systems and a detailed design of tomographic reconstruction algorithms,the device leverages the fast-pulse,multi-angle imaging capabilities of seven-focus cold cathode X-ray tube to achieve low-dose,high-resolution,three-dimensional imaging.The system operates at an anode voltage of 70 kV and anode current of 7 mA per focal spot.By sequentially activating the seven focal spots,it captures images from multiple angles within 1 second,re-constructing them into high-resolution 3D images of dental tissues with layered display capability,providing signifi-cantly more detailed information than conventional 2D imaging.The performance evaluation demonstrates that the measured anode voltage is 69.6 kV,with an aluminum half-value layer thickness of 2.13 mm,suitable for dental di-agnosis.The device achieves a 16 lp·mm-1 line-pair resolution,and the low-contrast resolution testing clearly re-veals a 0.75 mm diameter hole,demonstrating high-definition imaging characteristics.In the three-dimensional ima-ging mode,each focal spot operates at an anode current of 7 mA,with a total imaging scanning time of approxi-mately 700 ms for all seven spots,enabling rapid imaging for clinical convenience.The air kerma-area product for a single three-dimensional imaging is 46.4 mGy·cm2,significantly lower than the radiation dose of conventional cone beam computed tomography(CBCT).The three-dimensional imaging tests observe dental cracks,indicating good clinical application value.

唐华平;潘劲松;陈志强;金鑫;张庆辉;方润泽;李国宇;谢德华

清华大学 工程物理系,北京 100084||新鸿电子有限公司,江苏 常州 213200新鸿电子有限公司,江苏 常州 213200清华大学 工程物理系,北京 100084新鸿电子有限公司,江苏 常州 213200新鸿电子有限公司,江苏 常州 213200新鸿电子有限公司,江苏 常州 213200新鸿电子有限公司,江苏 常州 213200新鸿电子有限公司,江苏 常州 213200

机械制造

CNT冷阴极分布式X射线管低剂量三维成像高分辨率

CNT cold cathodeMulti-beam X-ray tubeLow radiation dose3D imagingHigh resolution

《真空电子技术》 2026 (1)

28-34,7

国家自然科学基金资助项目(12327809)

10.16540/j.cnki.cn11-2485/tn.2026.01.04