首页|期刊导航|右江医学|基于EP09-A3方案的IGF-1检测一致性评价:化学发光免疫分析法与电化学发光免疫分析法

基于EP09-A3方案的IGF-1检测一致性评价:化学发光免疫分析法与电化学发光免疫分析法OA

Consistency evaluation of IGF-1 measurements based on EP09-A3 protocol:chemiluminescence immunoassay and electrochemiluminescence immunoassay

中文摘要英文摘要

目的 评估新产业MAGLUM X8 全自动化学发光免疫分析仪(化学发光法)与罗氏E601 全自动电化学发光免疫分析仪(电化学发光法)检测胰岛素样生长因子 1(IGF-1)结果的一致性,为临床检验结果互认提供依据.方法 依据美国临床实验室标准协会(CLSI)EP09-A3 方案,收集 2023 年 7 月至 11 月河池市人民医院儿内科 80 份患者新鲜血清样本.分别使用两种仪器对同一样本在同一日内重复检测 2 次,采用Passing-Bablok回归分析、Bland-Altman图及医学决定水平偏倚评估等方法进行一致性评价.结果 Passing-Bablok回归方程为y=0.954+1.007x,斜率 95%CI(0.985,1.023)包含1,截距95%CI(-1.884,2.945)包含0,提示两系统无显著系统或比例偏差;Bland-Alt-man分析显示,两系统百分比差异的 95%一致性界限(LoAs)为-9.22%至 10.24%,均在±12.5%的临床可接受范围内,且 97.50%(78/80)的样本值位于LoAs区间内;我国未成年人IGF-1 参考区间阈值处的预期偏倚均在±12.5%的临床可接受范围内.结论 新产业MAGLUM X8 与罗氏E601 检测IGF-1 的结果具有高度一致性,可满足临床检验需求,为诊疗决策提供可靠支持.

Objective To evaluate the consistency of insulin-like growth factor 1(IGF-1)test results between the new MAGLUM X8 fully automated chemiluminescence immunoassay analyzer(chemiluminescence method)and the Roche E601 fully automated electrochemiluminescence immunoassay analyzer(electrochemiluminescence method),so as to provide a ba-sis for clinical test result recognition.Methods According to the Clinical Laboratory Standards Institute(CLSI)EP09-A3 protocol,80 fresh serum samples were collected from patients in the Department of Pediatrics at the People's Hospital of He-chi from July to November 2023.Two instruments were used to test the same sample twice on the same day,and consistency evaluation was performed using methods such as Passing-Bablok regression analysis,Bland-Altman plot,and medical deci-sion level bias assessment.Results The Passing-Bablok regression equation was y=0.954+1.007x.The 95%confidence interval(CI)for the slope(0.985,1.023)included 1,and the 95%CI for the intercept(-1.884,2.945)included 0,indicating no significant system atic or proportional bias between the two systems.Bland Altman analysis showed that the 95%consistency limit(LoAs)of the percentage difference between the two systems ranged from-9.22%to10.24%,both within the clinically acceptable range of±12.5%,and 97.50%(78/80)of the sample values were within the LoAs range;the expected bias at the IGF-1 reference interval threshold for minors in China was within a clinically acceptable range of±12.5%.Conclusion The results of testing IGF-1 using MAGLUM X8 and Roche E601 in the new industry are highly con-sistent,which can meet clinical testing needs and provide reliable support for diagnosis and treatment decisions.

吴鸿桃;孙莹莹;蓝晨;韦帅;陈高;覃猛

右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000右江民族医学院附属河池医院暨河池市人民医院医学检验科,广西 河池 547000

医药卫生

EP09-A3胰岛素样生长因子1化学发光免疫分析法电化学发光免疫分析法一致性

EP09-A3insulin-like growth factor 1(IGF-1)chemiluminescence immunoassayelectrochemilumines-cence immunoassayconsistency

《右江医学》 2026 (2)

152-158,7

深圳市医学检验分子诊断学重点实验室开放课题(ZDSYS20210623092001003)

10.3969/j.issn.1003-1383.2026.02.007

评论