高BMI人群CCTA中P3T个性化对比剂方案与标准方案的图像质量及对比剂用量对比研究OA

Comparative Study on Image Quality and Contrast Agent Dosage Between P3T Personalized Contrast Agent Protocol and Standard Protocol in CCTA for High BMI Population

中文摘要英文摘要

目的 探讨个体化患者对比剂注射技术(Personalized Patient Protocol Technology,P3T)在高体质量指数(Body Mass Index,BMI)(28~32 kg/m2)人群的冠状动脉CT血管造影(Coronary Computed Tomography Angiography,CCTA)临床应用价值.方法 前瞻性纳入2023年6-12月在首都医科大学附属北京友谊医院拟行CCTA扫描的患者65例,随机分成对照组(n=33)、P3T组(n=32).对照组根据不同BMI选择对应的对比剂注射流速.P3T组通过P3T软件根据患者的体重生成个性化的注射方案.比较2组的对比剂用量及注射流速、有效剂量(Effective Dose,ED)、图像质量的主观评价结果,以及图像质量的客观评价,包括主动脉根部及冠状动脉各段CT值、信噪比(Signal to Noise Ratio,SNR)和对比噪声比(Contrast to Noise Ratio,CNR).结果 P3T组对比剂用量和注射流速均低于对照组,差异有统计学意义(P<0.05).2组图像的主动脉根部及冠状动脉各段CT值、SNR、CNR和ED差异均无统计学意义(P>0.05).2组图像主观评分均≥3分,均能满足诊断需求,且一致性极好(Kappa=0.81).结论 高BMI人群中应用P3T技术能在保证图像质量的前提下减少对比剂用量,降低注射流速,实现注射方案个性化和检查的一致性,在临床上具有可行性.

Objective To investigate the clinical application value of personalized patient protocol technology(P3T)in coronary CT angiography(CCTA)for populations with high body mass index(BMI)(28-32 kg/m2).Methods A total of 65 patients scheduled for CCTA scanning in Capital Medical University Affiliated Beijing Friendship Hospital from June to December 2023 were prospectively enrolled,and randomly divided into a control group(n=33)and a P3T group(n=32).The control group selected the corresponding contrast agent injection flow rate according to different BMI values.The P3T group generated personalized injection protocols based on patients'body weight through the P3T software.The two groups were compared in terms of contrast agent dosage,injection flow rate,effective dose(ED),results of subjective evaluation of image quality,and objective evaluation of image quality[including CT values of the aortic root and each segment of the coronary artery,signal to noise ratio(SNR),and contrast to noise ratio(CNR)].Results The contrast agent dosage and injection flow rate in the P3T group were significantly lower than those in the control group(P<0.05).There were no statistically significant differences between the two groups in CT values of the aortic root and each segment of the coronary artery,SNR,CNR,and ED(P>0.05).The subjective scores of images in both groups were≥3 points,which could meet diagnostic needs,with good consistency(Kappa=0.81).Conclusion The application of P3T technology in high BMI populations can reduce contrast agent dosage and injection flow rate while ensuring image quality,realize the personalization of injection protocols and consistency of examinations,and is clinically feasible.

王宇鹏;张立新;吴柯薇;袁颖;李妍;胡艳军

首都医科大学附属北京友谊医院 放射科,北京 100050首都医科大学附属北京友谊医院 放射科,北京 100050首都医科大学附属北京友谊医院 放射科,北京 100050首都医科大学附属北京友谊医院 放射科,北京 100050首都医科大学附属北京友谊医院 放射科,北京 100050首都医科大学附属北京友谊医院 放射科,北京 100050

医药卫生

个体化患者对比剂注射技术(P3T)冠状动脉CT血管造影(CCTA)体质量指数(BMI)碘对比剂图像质量对比剂流速

personalized patient protocol technology(P3T)coronary CT angiography(CCTA)body mass index(BMI)iodinated contrast agentimage qualitycontrast agent flow rate

《中国医疗设备》 2025 (11)

133-137,149,6