多模型自适应统计迭代重建算法在颅脑CT扫描中最优权重组合的研究OA

Study on Optimal Weight Combination of Multi-model Adaptive Statistical Iterative Reconstruction in Cranial CT Scans

中文摘要英文摘要

目的 探讨在颅脑CT检查中,应用多模型自适应统计迭代重建算法(Adaptive Statistical Iterative Reconstruction-V,ASIR-V)前置和后置权重的最佳组合.方法 将90例颅脑CT平扫的患者应用随机数表法分为A组、B组和C组,每组30例,3组扫描前分别预设前置20%、30%和40%ASIR-V权重;每组扫描完成后均采用后置50%、60%、70%及80%ASIR-V权重进行图像重建,分别得到A5~A8亚组、B5~B8亚组和C5~C8亚组.分别记录3组扫描辐射剂量的剂量长度乘积、CT容积剂量指数,计算有效辐射剂量(Effective Dose,ED)及各组图像的信噪比(Signal to Noise Ratio,SNR)、对比噪声比(Contrast to Noise Ratio,CNR).由2名影像诊断医师采取双盲法应用5分制对图像进行主观评价.结果 客观评价方面,ED值随前置ASIR-V权重的增高而降低,两者呈线性负相关,ED值的回归方程为y=-0.33x+2.71(r2=0.918,P<0.001);A5~A8、B5~B8和C5~C8各亚组的SNR值、CNR值均随后置ASIR-V权重的增加而增加,差异具有统计学意义(P<0.001).主观评价方面,3组评分一致呈现先升高后降低的趋势,A6、B7和C7亚组分别在同一前置迭代水平下获得最高评分,而B7亚组的主观评分值在所有亚组中最高.结论 在噪声指数为3且管电压为120 kV的条件下,应用前置30%结合后置70%ASIR-V权重进行颅脑CT检查,可以降低噪声,提高图像质量,显著降低患者所接受的辐射剂量.

Objective To explore the optimal combination of pre-and post-weights for adaptive statistical iterative reconstruction-V(ASIR-V)in cranial computed tomography(CT)examinations.Methods Ninety patients undergoing cranial CT plain scans were divided into groups A,B,and C(30 cases each)using the random number table method.Pre-ASIR-V weights of 20%,30%,and 40%were preset for groups A,B,and C before scanning,respectively.After scanning,image reconstruction was performed with post-ASIR-V weights of 50%,60%,70%,and 80%,yielding subgroups A5-A8,B5-B8,and C5-C8.Radiation dose parameters(DLP,CTDIvol),effective dose(ED),signal-to-noise ratio(SNR),and contrast-to-noise ratio(CNR)were recorded.Two radiologists conducted subjective image evaluation using a 5-point scale in a double-blinded manner.Results In objective evaluation,ED decreased linearly with increasing pre-ASIR-V weight(regression equation:y=-0.33x+2.71,r2=0.918,P<0.001).SNR and CNR in all subgroups increased with higher post-ASIR-V weights(P<0.001).Subjective scores showed a trend of first increasing then decreasing;subgroups A6,B7,and C7 had the highest scores at their respective pre-ASIR-V levels,with subgroup B7 scoring the highest overall.Conclusion At a noise index of 3 and tube voltage of 120 kV,using 30%pre-ASIR-V and 70%post-ASIR-V weights in cranial CT reduces noise,improves image quality,and significantly lowers patient radiation dose.

朱超楠;田俊;孙文路;李前威;万婷婷;孙思阳

南京医科大学第二附属医院 放射影像科,江苏 南京 210011南京医科大学第二附属医院 放射影像科,江苏 南京 210011南京医科大学第二附属医院 放射影像科,江苏 南京 210011南京医科大学第二附属医院 放射影像科,江苏 南京 210011南京医科大学第二附属医院 放射影像科,江苏 南京 210011南京医科大学第二附属医院 放射影像科,江苏 南京 210011

医药卫生

CT颅脑CT自适应统计迭代重建算法(ASIR-V)辐射剂量图像质量

computed tomography(CT)cranial CTadaptive statistical iterative reconstruction-V(ASIR-V)radiation doseimage quality

《中国医疗设备》 2025 (9)

134-138,173,6

10.3969/j.issn.1674-1633.20241310