体部立体定向放射治疗摆位误差分布预测模型的构建OA
Construction of the Prediction Model of Setup Errors Distribution in Stereotactic Body Radiation Therapy
目的 构建立体定向体部放射治疗(Stereotactic Body Radiation Therapy,SBRT)患者摆位误差分布预测模型.方法 选取2019年1月至2022年12月在徐州医科大学附属医院肿瘤放射科使用美国瓦里安Vital Beam(SN3546)医用直线加速器治疗的45例SBRT患者的摆位误差数据为研究对象,采用变分推断贝叶斯高斯混合模型构建误差分布预测模型,求解模型参数.结果 平移方向摆位误差主要向3个中心(μ1~μ3)方向集中,在前后(-3.45~4.84 mm)和头脚方向(0.22~5.67 mm)平均偏移量较大,左右方向(0.14~0.79 mm)平均偏移较小.摆位误差向μ3中心点偏移可能性最大,误差中心概率为0.54,向μ1偏移可能性最小,误差中心概率为0.20,平移方向方差最大值为1.72 mm;旋转方向摆位误差主要向4个中心点(μ1~μ4)方向集中,在倾斜(-0.50°~0.15°)和旋转方向(-0.52°~-0.10°)平均偏移量相对自转(0.03°~0.25°)方向较大.摆位误差向μ3中心点偏移可能性最大,误差中心概率为0.69,其余3个中心点偏移可能性相差不大.与平移方向相比,旋转方向各中心方差值除μ2中心点数据(-0.52°)外数值均较大,说明平移方向数据更加聚集.结论 变分推断贝叶斯高斯混合模型可以对摆位误差数据进行定量描述和预测分析,为探寻SBRT摆位误差分布规律提供参考.
Objective To construct a prediction model for the setup error distribution of patients undergoing stereotactic body radiation therapy(SBRT).Methods The setup error data of 45 patients with SBRT who were treated with the Varian Vital Beam(SN3546)medical linear accelerator in the Department of Oncology Radiology of the Affiliated Hospital of Xuzhou Medical University from January 2019 to December 2022 were selected as the research objects.The error distribution prediction model was constructed by using the variational inference Bayesian Gaussian mixture model to solve the model parameters.Results The setup errors in the translation direction mainly concentrated in the three centers(μ1 to μ3)directions.The average offsets were relatively large in the front-back(-3.45 to 4.84 mm)and head-foot directions(0.22 to 5.67 mm),while the average offsets were relatively small in the left-right direction(0.14 to 0.79 mm).The possibility of offset of the setup error towards the center point of μ3 was the greatest,and the probability of the error center was 0.54.The possibility of offset towards μ1 was the smallest,and the probability of the error center was 0.20.The maximum variance in the translation direction was 1.72 mm.The setup errors in the rotation direction mainly concentrated in the directions of the four center points(μ1 to μ4).The average offsets in the tilt direction(-0.50° to 0.15°)and the rotation direction(-0.52° to-0.10°)were relatively larger than those in the rotation direction(0.03° to 0.25°).The possibility of the setup error offsetting towards the μ3 center point was the greatest,and the probability of the error center was 0.69.The offset possibilities of the remaining three center points were not much different.Compared with the translation direction,the variance values of each center in the rotation direction were larger except for the data of the μ2 center point(-0.52°),indicating that the data in the translation direction was more concentrated.Conclusion The variational inference Bayesian Gaussian mixture model can quantitatively describe and predict the setup error data,providing a reference for exploring the distribution regularity of SBRT setup errors.
李亮;陈鑫;范雪梅;薛冰;刘莹;章龙珍
徐州医科大学附属医院 放疗科,江苏 徐州 221006徐州医科大学 医学影像学院,江苏 徐州 221004徐州医科大学附属医院 放疗科,江苏 徐州 221006徐州医科大学附属医院 放疗科,江苏 徐州 221006徐州医科大学附属医院 放疗科,江苏 徐州 221006徐州医科大学附属医院 放疗科,江苏 徐州 221006
医药卫生
贝叶斯高斯混合模型摆位误差立体定向体部放射治疗(SBRT)预测模型锥形束CT图像引导
Bayesian Gaussian mixture modelsetup errorstereotactic body radiation therapy(SBRT)predictive modelcone-beam CTimage guidance
《中国医疗设备》 2025 (7)
54-60,7
徐州市引进临床医学专家团队项目(2019TD003)徐州医科大学附属医院院级科研项目(2022ZL19)徐州医科大学科技园创新创业项目(CXCYZX2022003).
评论